Prototipe Deteksi PEN dalam Tubuh Menggunakan B-Mode Ultrasonic Scanning
Abstract
Internal fixation is a method of healing fractures by connecting them using metal called pen. Pen metal in the body isn’t permanent, when the bone was strong the pen must be removed to avoid infection. In finding the pen, X-Ray are usually used. But X-Ray has bad effect, so not everyone is allowed to use it. The solution given by make prototype to detect pen location using B-Mode ultrasonic scanning method, it uses 5MHz single transducer
ultrasonic. Objects measurement is the animal bones connected with metals placed in the jelly with different heights. From the two transducers, the concave surface transducer has a better result and smaller percentage error than the flat surface transducer. The concave surface transducer has a focal length area 1.94cm above the transducer surface. The maximum slope angle of the object is 20o, which more than that makes the echo signal unreadable the object.
The measurement results of objects displayed in the B-Mode 1 dimensional on PC that displays the structure of the bottom surface of an object, based on comparison between the actual object and the B-Mode display an error less than 6%, meaning B-Mode can represent the surface structure bottom of the actual object.
Keywords
Full Text:
pdfReferences
https://www.klikdokter.com/tanya-dokter/read/2708883/pengaruh-x-ray, Diakses pada 2 Agustus 2018.
D. Segal , E. Palmanovich, A. Faour, E. Marom, V. Feldman, E. Yaacobi, O. Slevin, B. K. dan Y. S. Brin, “Routine early post-operative X-ray following internal fixation of intertrochanteric femoral fractures is unjustified: a quality improvement study”, Journal of Orthopaedic Surgery and Research, Vol. 13, No. 1, pp. 1-6, Juli 2018.
O. Lubovsky, M. Liebergall, Y. Mattan, Y. Weil, R. Mosheiff , “Early diagnosis of occult hip fractures MRI versus CT scan”, Injury, Vol. 36, No. 6, pp. 788-792, Juni 2005.
http://siui-ndt.ru/assets/files/UT-Probe.pdf, Diakses pada 25 Juli 2018.
L. R. Galván, R. R. Ruiz, dan E. S. Gómez, “C-Mode Ultrasonic System Based in Piezoelectric Ceramics and PVDF Polymers for Several Acoustic Impedances”, Electrical and Electronics Engineering, pp.1-4, Desember 2006.
http://teachersinstitute.yale.edu/curriculum/units/1983/7/83.07.05.x.html, Diakses pada 30 juli 2018.
https://www.olympus-ims.com/en/ndt-tutorials/thickness-gage/appendices-velocities/ diakses pada 2 Agustus 2018
J. Blahuta, T. Soukup, P. Čermák, dan M. Večerek, “Computer analysis and evaluation of ultrasound B-images universally usabledue to principle of B-MODE and its displaying in grayscale”, Communications, Signal Processing, and their Applications, pp. 1-6, April 2015.
J. Aerssens, S. Boonen, G. Lowet, dan J. Dequeker, “Interspecies Differences in Bone Composition, Density, and Quality: Potential Implications for in Vivo Bone Research”, Endocrinology, Vol. 139, No. 2, pp. 663-670, Februari 1998.
A. Pearce, R. Richards, S. Milz, E. Schneider dan S. Pearce, “Animal Mode ls For Implant Biomaterial Research In Bone: A Review”, European Cells and Materials, Vol. 13, pp. 1-10, Maret 2007.
M. S. Taljanovic, M. D. Jones, J. T. Ruth, J. B. Benjamin, J. E. Sheppard, dan T. B. Hunter, “Fracture Fixation”, RadioGraphic, Vol. 23, No. 6, pp. 1569-1590, November 2003.
A. I. Gunawan, N. Hozumia, K. Takahashia, S. Yoshidab, Y. Saijoc, K. Kobayashid, dan S. Yamamotoe, “Numerical analysis of acoustic impedance microscope utilizing acoustic lens transducer to examine cultured cells”, Ultrasonics, Vol. 63, pp. 102-110, Desember 2015.
A. I. Chen, M. L. Balter, M. I. Chen, D. Gross, S. K. Alam, T. J.Maguire, dan M. L.Yarmush, “Multilayered tissue mimicking skin and vessel phantoms with tunable mechanical, optical, and acoustic properties”, Medical Physics, Vol. 43, No. 6, pp. 3117-3131, Juni 2016.
J. R. Cook, R. R. Bouchard, dan S. Y. Emelianov, “Tissue-mimicking phantoms for photoacoustic and ultrasonic imaging”, Biomedical Optics Express, Vol. 2, No. 11, pp. 3193-3206, November 2011.
E. B. Prastika, A. I. Gunawan, B. S. B. Dewantara, C. E. Prianto, dan N. Hozumi, “The Enhancement of 3 MHz Ultrasonic Echo Signal for Conversion Curve Development for Acoustic Impedance Estimation by Using Wavelet Transform,” EMITTER International Journal of Engineering Technology, Vol. 6, No. 1, pp. 105-123, June 2018.
T. Namas, dan M. Dogruel, “A Feasible and Accurate Technique for Determining the Time-of-Flight in Ultrasonic Distance Measurements”, Electronics in Marine, Vol. 2, pp. 337-340, September 2008.
DOI: https://doi.org/10.17529/jre.v15i2.13893
Article Metrics
Abstract view : 0 timespdf - 0 times
Refbacks
- There are currently no refbacks.
View My Stats
Jurnal Rekayasa Elektrika (JRE) is published under license of Creative Commons Attribution-ShareAlike 4.0 International License.