Development of Solid State Reference Electrode Based on Polypyrrole (PPy) and Screen Printed Electrode (SPE)

Sagir Alva, Mas’ud Asadullah, Vivi Friliandita, Deni Shidqi Khaerudini

Abstract


In this study, a Polypyrrole doped KCl/KCl saturated (PPy-Cl/KCl) reference electrode had been successfully developed. In this development process, the PPy-Cl/KCl reference electrode that had been provided consisted of two stages of refinement, specifically the PPy-Cl/KCl 1 reference electrode which became the PPy-Cl/KCl 2 reference electrode. Furthermore, the PPy-Cl/KCl 2 SRE was made into 3 types with cycle variations of 10, 15, and 20 cycles and the mV response value test and Scanning Electron Microscopy (SEM) test were carried out. Regarding the variation results, PPy-Cl/KCl 2 SRE was obtained with 15 cycles as the best PPy-Cl/KCl 2 reference electrode with ∆mV value of 3.43 mV. Subsequently, the PPy-Cl/KCl 2 SRE with 15 cycles was involved in the characterization test using cyclic voltammetry (CV) testing, response testing vs Cl sensors and drift testing. In the CV testing, the PPy-Cl/KCl 2 SRE with 15 cycles indicated a voltammogram graphic pattern similar to the Ag/AgCl reference electrode as a commercial reference electrode. Moreover, in the response testing vs the Cl sensor, the PPy-Cl/KCl 2 SRE with 15 cycles provided a Nernstian number value of -54.5 mV/s with a testing range of 0.1-104 M, and in the PPy-Cl/SRE drift testing, KCl 2 with 15 cycles indicated fairly good stability, particularly with a drift value of 1 mV/minute which was achieved after the conditioning process after 3.5 minutes.


Keywords


Polypyrrole, solid-state, reference electrode, screen-printed electrode, SPE

Full Text:

PDF

References


Criscuolo,F., Galfione,M., Carrara,S., Micheli,G,D., (201s9), ”All-Solid State Reference Electrodes for analytical applications”, IEEE, Otranto, Italy DOI: 10.1109/IWASI.2019.8791382.

Dewi, M, S., Alva, S., Jamil, Q, A, W., (2021), “Development And Characterization Of Solid Cu/Cuso4 Reference Electrodes”, International Journal of Innovation in Mechanical Engineering & Advanced Materials (IJIMEAM), Vol.3 (No.1). 2021. pp. 17-25.

Pernaut, J.M., and Reynolds, J.R. 2000,“Use of Conducting Electroactive Polymers for Drug Delivery and Sensing of Bioactive Molecules. A Redox Chemistry Approach”J. Phys. Chem. B, 104, 17, DOI: 10.1021/jp994274o

Alva,S., Aziz,A., Syono,M.I., Sebayang,D., (2017), “Development of Solid-state Reference Electrode Based on Sodium Polyanethol Sulfonate Immobilised on Cellulose Acetate,”Journal of Physical Science, Vol. 28(2), hal :161–179.

Alva,S., Suherman,R., Friliandita,V., Khaerudini,D.S., Majlan,E.H., Aziz,A.S.A,(2020),”Preliminary Study of Poly(Tetrahydrofurfuryl Acrylate) Thin Film as a Potential Material of Ion Selective Electrodes: The Case of Nitrate Ion-Selective Electrode, Indones. J. Chem., 20 (3), hal: 645 - 654.

Alva,S., Ardiyansyah,D., Khaerudini,D.S., Suherman,R.,(2019c),”Solid-State Reference Electrode Based on Thin-Films of Tetrahydrofurfuryl Acrylate (pTHFA) Photopolymer,“Journal of The Electrochemical Society, 166 (8), hal: B598-B603.

Gunawan, B., &Azhari, C. D. (2010). KarakterisasiSpektofotometri I R dan Scanning Electron Microscopy (SEM) Sensor Gas dariBahanPolimer Poly Ethelyn Glycol (PEG). JurnalSainsdanTeknologi, 3(2

Elgrishi,N., Rountree,K.J., McCarthy, B.D., Eric S. Rountree,E.S., Thomas T. Eisenhart,T.T., Jillian L. Dempsey,J.L., (2018), “A Practical Beginner’s Guide to Cyclic Voltammetry,” J. Chem. Educ. 95, hal: 197−206.

S. Alva, A. S. Binti Abdul Aziz, M. I. Bin Syono, and W. A. Bin Wan Jamil, “Ag/AgCl reference electrode based on thin film of arabic gum membrane,” Indones. J. Chem., vol. 18, no. 3, pp. 479–485, 2018, doi: 10.22146/ijc.28859.

Khaerudini,D.S, Rahman,F., Alva,S., (2020),”Optimization strategy of Ag/AgCl thin film electrodes approached by chlorination process for electrochemical response materials,” Materials Chemistry and Physics, Volume 240, hal: 122294.

K.S.Ying, L.YHeng, N.I.Hassanan, S.A.Hasbullah,” A New and All-SolidState Potentiometric Aluminium Ion Sensor for Water Analysis,” Sensors vol. 20, pp. 6898, 2020




DOI: https://doi.org/10.24815/jocarbazon.v2i1.36417

Article Metrics

Abstract view : 0 times
PDF - 0 times

Refbacks

  • There are currently no refbacks.


Copyright (c) 2023 Journal of Carbazon

E-ISSN: 2988-5744

 

This work is licensed under a Creative Commons Attribution 4.0 International License (CC BY 4.0).